[1]蒋侃凌,田 维,贾 健.TiRobot手术机器人辅助经皮骶髂螺钉固定治疗骨盆后环不稳定损伤[J].天津医科大学学报,2017,23(03):247-251.
 JIANG Kan-ling,TIAN Wei,JIA Jian.TiRobot surgical robotic navigation and location system assisted percutaneous iliosacral screws for treatment of unstable pelvic posterior ring injuries[J].Journal of Tianjin Medical University,2017,23(03):247-251.
点击复制

TiRobot手术机器人辅助经皮骶髂螺钉固定治疗骨盆后环不稳定损伤(PDF)
分享到:

《天津医科大学学报》[ISSN:1006-8147/CN:12-1259/R]

卷:
23卷
期数:
2017年03期
页码:
247-251
栏目:
临床医学
出版日期:
2017-05-18

文章信息/Info

Title:
TiRobot surgical robotic navigation and location system assisted percutaneous iliosacral screws for treatment of unstable pelvic posterior ring injuries
文章编号:
1006-8147(2017)03-0247-05
作者:
蒋侃凌1 田 维2 贾 健 2
(1.天津医科大学研究生院,天津300070; 2.天津市天津医院骨盆科,天津300211)
Author(s):
JIANG Kan-ling1TIAN Wei2JIA Jian2
?(1. Graduate School , Tianjin Medical University,Tianjin 300070,China;2.Department of Pelvis,Tianjin Hospital of Tianjin,Tianjin 300211,China)
关键词:
计算机辅助骶髂关节内固定器骨盆
Keywords:
computer-assistedsacroiliacjoint internal fixatorspelvis
分类号:
R683.3
DOI:
-
文献标志码:
A
摘要:
目的:探讨运用TiRobot外科手术机器人导航定位系统辅助经皮骶髂螺钉固定治疗骨盆后环不稳定损伤患者的安全性和可行性。方法:回顾性分析运用TiRobot手术机器人经皮置入骶髂螺钉治疗骨盆后环不稳定损伤的20例病例,其中男12例,女性8例;年龄31~58岁;依据AO/OTA损伤分型:B型损伤14例,C型损伤6例。结果:20例患者共置入34枚螺钉,术中骶髂螺钉设计规划置入时间10~28 min,术中透视时间:6~12 s,手术出血量10~120mL。术后复查所有患者螺钉置入位置满意,无1例螺钉错位、切出骨质或切入骶孔及骶管,无切口感染、手术导致的血管神经并发症。术后复查采用Majeed评分标准,优12例,良5例,可3例,总体优良率85%。结论:运用Tirobot外科手术机器人导航定位系统经皮骶髂螺钉固定治疗骨盆后环不稳定性损伤患者是提高骨盆微创手术效果安全有效的手段。
Abstract:
Objective To evaluate effectsand safetyof using TiRobot surgical robotic navigation and location system assisted percutaneous iliosacral screws for treatment of unstable pelvic posterior ring injuries.Methods: Twenty patients with unstable pelvic postierior ring injuries,who were treated using new surgical robotic insertion percutaneousiliosacral screws, were analyzed retrospectively.There were 12 men cases,8 female cases,age:31~58 years.According to AO/OTA classification, there were 14 cases of type B,6 cases of type C. Results: there were total of 34 iliosacral screws inserted in 20 patients.In the operation, the insertion time of iliosacral screws was 10-28min,fluoroscopy time was 6-12sand the blood loss was 10-120mL.The review of screw position was satisfactory in all the 20 cases.None of the screws were malpositioned,penetrated bone cortex or entranced the sacral cannal and sacral foramen after surgery. There were no wound infection and neurovascular injury in 20cases. Based on the Majeed functional scoring,the results were excellent in 12 cases,good in 5 cases and plain in 3 cases. The excellent and good rates was 85%.Conclusion:UsingTiRobot surgical robotic navigation and location system insertion percutaneousiliosacral screws for treating unstable pelvic postierior ring injuries may be an effective and safe to improve pelvic invasive surgical results.

参考文献/References:

[1]Giannoudis P V, Grotz M R, Tzioupis C, et al.Prevalence of pelvic fractures, associated injuries, and mortality: The united kingdom perspective[J].J Trauma,2007,63(4):875

[2]Denis F, Davis S, Comfort T.Sacralfractures:an important problem.Retrospective analysis of 236 cases[J].ClinOrthopRelat Res,1988(227):67

[3]Dudda M, Hoffmann M, Schildhauer T A.Sacrum fractures and lumbopelvic instabilities in pelvic ring injuries: classification and biomechanical aspects[J].Unfallchirurg,2013,116(11):972

[4]Lindsay A.Tornetta P 3rd,Diwan A, et al. Is closed reduction andPercutaneousfixation of unstable posterior ring injuries as accurate as open reduction and internal fixation[J].JOrthop Trauma,2016,30(1):29

[5]Barei D P, Shafer B L, Beingessner D M, et al.The impact of open reduction internal fixation on acute pain management in unstable pelvic ring injuries[J].J Trauma,2010,68(4):949

[6]Pavelka T, Salá?ek M, WeisováD.Complications associated with surgical treatment of pelvic ring fractures[J].ActaChirOrthopTraumatolCech,2013,80(3):208

[7]Mardanpour K, RahbarM.The outcome of surgically treated traumatic unstable pelvic fractures by open reduction and internal fixation[J].J Inj Violence Res,2013,5(2):77
[8]Matityahu A, Kahler D, Krettek C, et al.Three-dimensional navigation is more accurate thantwo-dimensional navigation or conventional fluoroscopy for percutaneoussacroiliac screw fixation in the dysmorphicsacrum:a randomized multicenterstudy[J].J Orthop Trauma,2014,28(12):707

[9]邓宁,吴伟坚,梁国穗.机器人和计算机辅助骨科手术[J].中华创伤骨科杂志,2005,7(7):620

positioning ofsacroiliac screws in 88 patients[J].J Trauma,1994,53(1):51
[10]van den Bosch EW,vanZwienenCM,vanVugtAB.Fluoroscopic positioning ofsacroiliac screws in 88 patients[J].J Trauma,1994,53(1):51

[11]Insche A F, Giannoudis P V, Smith R M.Fluoroscopy-basedmultiplanar image guidance for insertion of sacroiliac screws[J].ClinOrthopRelat Res,2002,395(Feb):135

[12]Templeman D, Schmidt A, Freese J, et al.Proximity of iliosacral screws to neurovascular structures after internal fixation[J].ClinOrthopRelat Res,1996(329):194

[13]Verbeek J, Hermans E, van Vugt A, et al.Correct positioning of percutaneous iliosacral screws with Computer-Navigated versus fluoroscopically guided surgery in traumatic pelvic ring fractures[J].J Orthop Trauma,2016,30(6):331
[14]Zwingmann J, Konrad G, Mehlhorn A T, et al.Percutaneousiliosacral screw insertion:malpositioning and revision rate of screws withregards to application technique(navigatedvsConventional)[J].J Trauma, 2010, 69(6):1501
[15]St?ckle U, Schaser K, K?nigB.Image guidance in pelvic and acetabular surgery--expectations, success and limitations[J].Injury,2007,38(4):450

[16]Zwingmann J H, Hauschild O, Bode G, et al.Malposition and revision rates of different imaging modalities for percutaneous iliosacral screw fixation following pelvic fractures: asystematic review and meta-analysis[J]. Arch Orthop Trauma Surg, 2013,133(9):1257
[17]Arand M, Kinzl L, GebhardF.Computer-guidance in percutaneous screw stabilization of the iliosacraljoint[J].ClinOrthopRelat Res,2004(422):201

[18]Theologis A A, Burch S, PekmezciM.Placement of iliosacral screws using 3D image-guided (O-Arm) technology and Stealth Navigation: comparison with traditional fluoroscopy[J].Bone Joint J,2016,98 (5):696

[19]Coste C, Asloum Y, Marcheix P S, et al.Percutaneousiliosacral screw fixation inunstable pelvic ring lesions:the interest of O-ARM CT-guidednavigation[J].doi:10.1016/j.otsr,2013,99(4Suppl):S273
[20]苏永刚,王军强,王豫,等.双平面骨科机器人辅助骶髂关节螺钉置入系统精度测试及尸体应用的研究[J].中华创伤骨科杂志,2008,10(5):460

[21]苏永刚,王军强,刘文勇,等.双平面骨科机器人系统辅助骶髂关节螺钉置入的实验研究[J].中华创伤骨科杂志,2006,8(1):45

[22]Webb J E, Regev G J, Garfin S R, et al.Navigation-assisted fluoroscopy in minimally invasive direct lateral interbody fusion: a cadaveric study[J].SAS J,2010,4(4):115

[23]Kim J W, Oh C W, Oh J K, et al.Percutaneousiliosacral screwing in pelvic ring injury using three-dimensional fluoroscopy[J].J OrthopSci,2013,18(1):87

[24]Zwingmann J, Konrad G, Kotter E, et al. Computer-navigated iliosacral screw insertion reduces malposition rate and radiation exposure[J].ClinOrthopRelat Res,2009,467(7):1833

[25]王军强,刘达,赵春鹏,等.三维导航结合机械臂手术系统经皮螺钉固定骨盆环的实验研究[J].中华外科杂志,2012,50(6):555

?

相似文献/References:

备注/Memo

备注/Memo:
作者简介蒋侃凌(1991-),男,硕士在读,研究方向:创伤骨盆骨折;通信作者:贾健,E-mail:professorjiajian@126.com