[1]王欣宇,李峰坦.自适应统计迭代重建算法在头部低剂量CT扫描中的应用[J].天津医科大学学报,2015,21(02):161-163.
 WANG Xin-yu,LI Feng-tan.Application of adaptive statistical iterative reconstruction algorithm in low-dose head CT[J].Journal of Tianjin Medical University,2015,21(02):161-163.
点击复制

自适应统计迭代重建算法在头部低剂量CT扫描中的应用(PDF)
分享到:

《天津医科大学学报》[ISSN:1006-8147/CN:12-1259/R]

卷:
21卷
期数:
2015年02期
页码:
161-163
栏目:
临床医学
出版日期:
2015-03-20

文章信息/Info

Title:
Application of adaptive statistical iterative reconstruction algorithm in low-dose head CT
文章编号:
1006-8147(2015)02-0161-03
作者:
王欣宇李峰坦
(天津医科大学总医院放射科,天津 300070)
Author(s):
WANG Xin-yuLI Feng-tan
 (Department of Radiology, General Hospital,Tianjin Medical University, Tianjin 300070,China)
关键词:
宝石CT低剂量CT扫描自适应统计迭代重建技术图像质量
Keywords:
low-dose computed tomography adaptive statistical iterative reconstructionimage quality
分类号:
R445
DOI:
-
文献标志码:
A
摘要:
目的:使用自适应统计迭代重建技术( ASiR)算法,在宝石CT进行头颅扫描,探讨头部低剂量扫描的可行性。方法:患者90例行头部CT检查,随机分为A、B、C 3组(每组30例)。3组病例扫描参数相同,A组单纯行滤波反投影算法( FBP)重建;B组行FBP重建后,打开减少剂量因子DR,设定为50%;C组行ASiR重建后,打开减少剂量因子DR,同样设定为50%,其余扫描条件3组相同。对3组图像,分别测量空气及侧脑室前脚脑脊液CT值的标准差,取均数作为图像噪声,由3位医师对3组数据进行评分。对3组数据噪声值及评分分别进行方差分析和秩和检验。结果:A组的容积CT剂量指数(CTDIvol)为45.79 mGy;B组为22.75 mGy;C组为 22.75 mGy,B、C组较A组降低了50.31%。C组图像噪声(1.96±0.52)显著低于B组(2.92±0.67)(p<0.001),与A组(2.12±0.53)无显著性差异(p=0.294)。3组图像数据均能满足诊断要求,B组图像评分较低,A组和C组无显著性差异。结论:在颅脑CT扫描中使用ASiR算法,保证图像质量的同时可以大幅降低辐射剂量。
Abstract:
Objective: To explore the feasibility of low-dose scan in head CT using adaptive statistical iterative reconstruction (ASiR) algorithm in gemstone CT. Methods: Ninety patients who received head CT scans were included, were randomly divided into three groups (n = 30). Standard dose scan group and low dose scan group were treated with the same tube current 300 mA and 260 kV, respectively. After applying dose reduction factor (DR) 50% in B and C group, no differences were found in other scanning parameters. Standard dose scans were reconstructed with filter back projection (FBP) (group A) while low dose scans were performed with FBP (group B) and 50% ASiR (group C). For all 3 groups, s values of air and cerebrospinal fluid in anterior horns of lateral ventricles were measured, the mean value was taken as image noises, and all data were scored by 3 radiologists. Noise values and scores of 3 groups were compared by ANOVA and rank sum test. Results: CTDIvol was lowered by 50.31% in low dose group as compared to standard dose group. Noise measured in group C (1.96 ± 0.52) was significantly lower than group B (2.92 ± 0.67) (p <0.001). However, no significant difference (p = 0.294) was found between group A (2.12 ± 0.53)and group C. All three sets of data could meet the diagnostic requirements. Group B had a lower imaging score, but no significant difference between group A and group C was found. Conclusion: Applying ASiR algorithm in head CT scan can significantly reduce radiation dose while maintaining image quality and meeting diagnostic requirements.

参考文献/References:

[1]Smith P R, Peters T M, Bates. RHT:image Reconstruction from finite numbers of projections[J]. J Phys, 1973, 6: 361

[2]胡德纯.优质X线影像密度标准的监测及标准值[J].中外医用放射技术,1993,4(3):5

[3]Kristina T F, Amyk H, Silva A C. Reducing the radiation dose for CT colonography using adaptive statistical iterative reconstruction:a pilot study[J]. AJR ,2010,195(1):126

[4]Thibault J B, Sauer K D, Bouman C A, et al. A three-dimensional statistical approach to improved image quality for multislice helical CT[J]. Med Phys, 2007, 34(11): 4526

[5]Singh S, Kalra M K, Hsieh J, et al. Abdominal CT: comparison of adaptive statistical iterative and filtered back projection Reconstruction techniques[J]. Radiology, 2010, 257(2): 373

[6]杨尚文,何健,杨献峰,等.ASiR算法结合自动管电流调制技术在胸部低剂量CT中的应用研究[J].医学影像学杂志,2012,22(1):58

[7]Fujii K, Aoyama T, Koyama S, et al. Comparative evaluation of organ and effective doses for paediatric patients with those for adults in chest and abdominal CT examinations[J]. Br J Radiol, 2007, 80(956): 657

[8]Flicek K T, Hara A K, Silva A C, et al. Reducing the radiation dose for CT colonography using adaptive statistical iterative Reconstruction: A pilot study[J]. AJR Am J Roentgenol, 2010, 195(1): 126

[9]Maffei E, Martini C, Rossi A, et al. Diagnostic accuracy of second-generation dual-source computed tomography coronary angiography with iterative reconstructions: a real-world experience[J]. Radiol Med, 2012, 117(5): 725

[10]Leipsic J, Nguyen G, Brown J, et al. A prospective evaluation of dose reduction and image quality in chest CT using adaptive statistical iterative Reconstruction[J]. AJR Am J Roentgenol, 2010, 195(5): 1095

相似文献/References:

备注/Memo

备注/Memo:

基金项目 天津市卫生局科研基金资助项目(2010KZ106)

作者简介 王欣宇(1986-),男,初级技师,学士,研究方向:医学影像学;通信作者:李锋坦,E-mail :left9999@sina.com



更新日期/Last Update: 2015-03-24